Specs at a Glance

Process pressure:
0.8 to 11 bar (11.6 to 159.5 psi) absolute
Measurement range:
Application A
• pH: 1 to 12
Application B
• pH: 0 to 14
Process temperature:
Application A: –15 to 80 °C (5 to 176 °F)
Application B: 0 to 135 °C (32 to 275 °F)

Field of application

The Memosens CPS41E features a liquid KCl electrolyte and a ceramic junction, making it extremely suitable for harsh chemical applications, fast-changing media and liquids with low conductivity or high organic content. Its integrated Memosens 2.0 digital technology enables extended data storage and lab calibration which inputs in easier operation and more process uptime and provides the perfect basis for predictive maintenance. The non-contact signal transmission ensures integrity of your process.

Specifications

Product Family

  • Memosens

pH

Application

  • Media with very low conductivity or a high proportion of organic solvents or alcohol:
  • • Chemical industry
  • • Organic chemicals
  • • Power stations
  • • Laboratory measurements

Characteristic

  • Digital pH electrode for process engineering with ceramic junction
  • and KCl liquid electrolyte

Measurement range

  • Application A
  • • pH: 1 to 12
  • Application B
  • • pH: 0 to 14

Measuring principle

  • Liquid-KCl compact electrode with ceramic junction

Design

  • All shaft lengths with temperature sensor

Material

  • Sensor shaft: Glass to suit process
  • pH membrane glass: Type A and B
  • Metal lead: Ag/AgCl
  • Open aperture: Ceramic junction, zirconium dioxide
  • O-ring: FKM
  • Process coupling: PPS fiber-glass reinforced
  • Nameplate: Ceramic metal oxide

Dimension

  • Diameter: 12 mm (0.47 inch)
  • Shaft length: 120, 225, 360 and 425 mm
  • (4.72, 8.86, 14.17 and 16.73 inch)

Process temperature

  • Application A: –15 to 80 °C (5 to 176 °F)
  • Application B: 0 to 135 °C (32 to 275 °F)

Process pressure

  • 0.8 to 11 bar (11.6 to 159.5 psi) absolute

Temperature sensor

  • NTC 30K

Ex certification

  • With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
  • hazardous areas Zone 0, Zone 1 and Zone 2.

Connection

  • Inductive, digital connection head with Memosens 2.0 technology

Ingress protection

  • IP68

Additional certifications

  • Additional certifications

  • Chemical Industry
  • Organic Chemicals
  • Power Stations
  • Laboratory Measurements
show more